Abstract

A comparison CCAUV.V-S1 was organized to compare measurements of sinusoidal angular accelerations in the frequency range from 0.4 Hz to 1 kHz. This was a bilateral comparison between the Korea Research Institute of Standards and Science (KRISS), Republic of Korea, and the Physikalisch-Technische Bundesanstalt (PTB), Germany.Both NMIs applied laser interferometry in compliance with ISO 16063-15:2006 'Methods for the calibration of vibration and shock transducers—Part 15: Primary angular vibration calibration by laser interferometry'.The complex voltage sensitivity (magnitude and phase) of one angular transfer standard in the frequency range of 0.4 Hz to 100 Hz and the complex charge sensitivity (magnitude and phase) of one angular reference standard in the frequency range of 1 Hz to 1000 Hz were measured.Main text.To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the CCAUV, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call