Abstract

This study aims at investigating the possible employment for pre-treatment verifications in Intensity Modulated Radiation Therapy (IMRT) of a polycrystalline Chemical Vapour Deposited (pCVD) diamond bidimensional detector.The pCVD device, with an area of 2.5 × 2.5 cm2 (12 × 12 pixels), has been used to measure dose maps of a 10 MVRX prostatic IMRT field. Its response was compared both with a commercial bi-dimensional detector made with silicon and with Treatment Planning System (TPS) calculations.Measurement provided promising results on a map of 1.8 × 12.6 cm2. Absorbed doses measured along IMRT profiles by our device are consistent with the ones acquired with the commercial device and an overall good agreement with respect to the TPS was found for the diamond dosimeter.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.