Abstract
A biamperometric methodology for the sequential determination of thorium (Th) and uranium (U) was developed. In the sample solution containing Th and U, Th was first determined by complexometric titration based on the electrochemical behaviour of EDTA maintaining a potential of ≥200 mV between the twin Pt electrodes. This was followed by the redox titrimetric determination of U employing biamperometry to detect the end point. Prior to the determination of U, EDTA was destroyed by fuming with concentrated HClO4 to eliminate its interference in the U determination. The method was tested for the determination of Th and U in (Th, U) O2 samples containing 4 mg of Th and 2–8 mg of U, with precision and accuracy of better than 0.3 %.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.