Abstract

An inpatient endoscopy unit is a care hub for patients from throughout the hospital and can be the site of healthcare-associated infections (HAIs). Shared surfaces and other nonmedical devices (keyboards) have been increasingly recognized as sites of pathogen transmission. Beyond standard cleaning of high-touch target areas, we queried whether the addition of automated devices delivering low-intensity doses of ultraviolet (UV)-C radiation could further reduce bioburden in an academic endoscopy unit. Bioburden on previously identified high-touch/communal surfaces was measured before and after the installation of automated, low-intensity UV-light emitting devices (UV Angel) that passively monitor and disinfect targeted surfaces with Ultraviolet-C light (UV-C) light. High-touch sites (keyboards) had a baseline bacterial contamination of >80%, whereas individual procedure rooms and common areas had a >57% contamination rate. Following the implementation of automated UV-C light decontamination, bioburden was reduced on average by >91% at high-touch surfaces and within procedure rooms. Nonsterile hubs of patient care could serve as sites of "silent" HAI transmission. We have identified high-touch surfaces within an endoscopy unit that have a high bioburden of bacterial contamination and demonstrated that the installation of passive, automated UV-C light disinfection devices can reduce bioburden significantly, possibly mitigating HAI transmission between patients.