Abstract
An autoradiography system prototype (BETA view) is presented; it is aimed at quantitative dynamic analysis of radioactive labelled biological samples. The system is based on a GaAs pixel array detector, bump-bonded to a low threshold and single particle counting electronics (Medipix). The detector has an area of about 1 cm 2, segmented in 64×64 square pixels, 200 μm thick. Studies with gamma and beta emitters ( 210Pb, 241Am, 152Eu, 32P, 90Sr and 14C) allowed to identify optimal detector bias, to estimate detection efficiency and to measure system counting linearity. Two experiments showed the system capability to select radionuclides with different beta spectra and to perform a real time monitoring of biological phenomena.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.