Abstract

The Gamma Knife is used as a stereotactic tool for the conformal treatment of very small, complex-shape cranial lesions. The combination of planning software and treatment equipment enables a highly-precise conformal dose distribution and positioning. The purpose of the present study was to experimentally verify the precision actually achievable in case of extremely irregular, small target volumes. For this purpose, a complete treatment procedure was performed using a standard head phantom complemented with a specially developed insert that simulates an L-shaped lesion. The spatial precision of the irradiation was recorded by means of high-resolution film dosimetry using GafChromicTM films. The analysis of the films showed for the film in the center plane an excellent conformity of the 75% isodose line used to circumscribe the lesion. A very good agreement between planning and measurement resulted also for isodose lines residing outside of the target volume.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.