Abstract

The elements Li, Na, K, Rb, Cs, Ba, La and Tl, having their strongest atomic lines in the wavelength range from 500 to 1000 nm, are exited in a Grimm glow discharge lamp. The radiation is detected by a silicon photodiode array which shows a good spectral response in the mentioned region. Electrically non-conductive rock powders (NIM samples) serve as sample materials which contain the elements named above as trace elements only, in the majority of cases. The standard deviation which is essential for the intensity measurement can be determined by the measuring system and by fluctuations of the radiation, and it is investigated examplary in some analysis. The influence of the graininess gathering particular signification in the analysis of heterogeneous samples is shown, and the effect on precision and accuracy is proven experimentally and by means of Wilson's theory. Furthermore, calibration curves, limits of detection and the weight tractions of the elements Li, Rb, Ba and partially Na, K and La in rock samples are reported.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.