Abstract
The determination of the absorbed dose with ionisation chambers is mainly influenced by the energy dependence of the mass stopping power ratio of the phantom material water to the material air of the probe volume. Also the material and dimensions of the ionisation chamber, represented by the pertubation factor, give energy dependent corrections. For a set of standard ionisation chambers this dependence will be presented up to an energy of 45 MV.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.