Abstract

ABSTRACTThis paper deals with some important (but often neglected) details about the uncertainty of retention measurement in thin layer chromatography, the propagation of uncertainty during computing simple and more complex values from the retention data, ending in influence of the retention uncertainty onto the regression estimates during extrapolation and lipophilicity estimation. Theoretical considerations are tested on data from previous study. It can be concluded that when TLC spots are broad and the retention uncertainty exceeds about 0.02 of RF value, the uncertainty should be taken into the account in further computations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.