Abstract
This work investigates the effect of surface texturing on the coefficient of friction (COF) of tibial inserts (TIs), of total knee replacements (TKRs), using a servomechanical knee simulator operating under the conditions prescribed by ISO 14243-3:2014, which requires strict control of flexion/extension (FE), anterior/posterior displacement (AP), inward/outward rotation (IOR) movements, and load variation. Tests were carried out lubricating with fetal bovine serum solution with a protein concentration of 20 g/L at 37 ± 2 °C. TIs were manufactured by fused deposition modeling (FDM) using polylactic acid (PLA) because of the advantages associated with rapid prototyping and customization. The acquired COF data were analyzed using statistical parametric mapping (SPM) to determine how the surface of the TIs affects the COF, identifying ranges where significant differences occurred. Textured surfaces led to significant reductions in COF during the swing phase of the walking cycle.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.