Abstract

The MARVEL study demonstrated at a single time point that accelerometer (ACC)-based atrial sensing improves atrioventricular (AV) synchrony (AVS) in patients with AV block and a Micra pacemaker (Medtronic, Minneapolis, MN, USA). The purpose of the MARVEL Evolve substudy was to assess the performance over time. This prospective single-center study compared AVS and ACC signals at two visits ≥6months apart. Custom software was temporarily downloaded into the Micra at each visit and AVS was measured during 30min at rest. Nine patients from the MARVEL study were enrolled. The mean (±standard deviation) age was 82.3±6.0 years old, 67% were male, and a Micra was implanted for 6.0±6.4 months. High-degree AV block was present in four patients, whereas five with predominantly intrinsic conduction required intermittent pacing for bradycardia. The mean interval between visits was 7.1±0.6 months. Seven patients had normal sinus node function at both visits and were included in a paired analysis. Both ACC signal amplitude (visit 2-visit 1=1.4 mG; 95% confidence interval [CI] [-25.8 to 28.4 mG]; P=0.933) and AVS (visit 1: 90.8%, 95% CI [72.4, 97.4] and visit 2: 91.4%, 95% CI [63.8, 98.5]; P=0.740) remained stable. Three patients had spontaneous atrial tachycardia. During atrial fibrillation, no atrial contraction was detected or tracked. During atrial flutter, intermittent tracking resulted in a ventricular rate of 60±8beats per minute (bpm); there was no ventricular pacing>100bpm. ACC signals amplitude and performance of AVS pacing were stable over time. During atrial arrhythmias, the AV synchronous pacing mode behaved safely.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call