Abstract

To further improve analytical accuracy in clinical chemistry, proficiency testing needs amelioration in the quality of materials tested and in target value assignment. To obtain information on the actual state-of-the-art in the Lombardy region of Italy, and to examine the behavior of different types of control materials (fresh-frozen human sera and lyophilized materials), we developed the following experimental design. Two human serum pools and two lyophilized sera were distributed to 32 laboratories for determination of glucose, creatinine, cholesterol, sodium, potassium, and gamma-glutamyltransferase (gamma-GT). Each analyte was measured in triplicate on each of 3 days. Target values for the controls were obtained with Reference Methods. The results show a good intralaboratory precision for every component but some accuracy problems for glucose, creatinine, cholesterol, and gamma-GT. Lyophilized materials showed some commutability problems for glucose, electrolytes, and gamma-GT, mainly with dry chemistry technology.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.