Abstract
BackgroundPeripherally inserted central catheters (PICCs) are typically placed under fluoroscopy. We used a magnetic tracking and electrocardiogram-based tip confirmation system for insertion of a PICC insertion in a morbidly obese patient at the bedside.Case presentationA 53-year-old female with severe obesity (height, 160 cm; weight, 217 kg; BMI, 84.8 kg/m2) was admitted to the intensive care unit. Both bilateral, inguinal, and cervical regions were covered with an excess of adipose tissue, making it difficult to place a central venous line. Since transferring her to fluoroscopy seemed dangerous, a PICC was inserted using Sherlock 3CG® TCS at the bedside. Magnetic sensor guidance failed due to the thick subcutaneous tissue her precordium, but intracavity electrocardiography could direct the tip to an appropriate position.ConclusionWe experienced bedside insertion of a PICC into a patient with BMI of 84.8 kg/m2 patient using a Sherlock 3CG® TCS. Since the interaction between Sherlock 3CG® TCS and body habitus has not been investigated, further reports are needed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.