Abstract

In order to obtain power spectral information on the fetal heart rate in stages of pregnancy earlier than labor an algorithm has been developed to calculate the fetal heart rate on a beat-to-beat basis from Doppler ultrasound cardiotocographic signals. The algorithm was evaluated by comparing the calculated fetal heart rate with the heart rate determined from direct ECG signals measured with a scalp electrode. Heart rates were compared both in time and frequency domain. In the time domain the results achieved by both methods correlate well (correlation coefficient = 0.977 (p < 0.001)), in the frequency domain the results correlate even better (correlation coefficient = 0.991 (p < 0.001)). Based on these findings, it can be concluded that the developed algorithm provides a valuable tool for obtaining power spectral information on the fetal heart rate in stages of pregnancy earlier than labor.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.