BackgroundWe present two cases of unicompartmental knee arthroplasty (UKA) bearing failure in this report—one case of bearing dislocation and one case of bearing fracture. The causes of failure in both cases are evaluated in depth and recommendations are provided regarding intraoperative technique to reduce risk of bearing failure in mobile bearing UKAs.Case presentationIn the first case, intraoperative evidence of metallosis and chronic pain preceding the traumatic event may indicate that the patient had attenuation of her collateral ligaments that precipitated the instability event. In the second case, the relatively atraumatic nature of the bearing fracture-dislocation and intraoperative evidence of extensive poly wear suggest that the bearing fracture was likely due to a 3-mm bearing selection in the initial surgery.ConclusionsThis case report shows that late bearing in mobile bearing unicompartmental knee arthroplasty can often be a multifactorial event and treatment must address all the risk factors that led to bearing dislocation. Bearing fracture is a very rare complication associated with mobile bearing UKA and patients with thin polyethylene inserts are at risk for bearing fracture even in the absence of poly wear.

Highlights

  • We present two cases of unicompartmental knee arthroplasty (UKA) bearing failure in this report— one case of bearing dislocation and one case of bearing fracture

  • This case report shows that late bearing in mobile bearing unicompartmental knee arthroplasty can often be a multifactorial event and treatment must address all the risk factors that led to bearing dislocation

  • Bearing fracture is a very rare complication associated with mobile bearing UKA and patients with thin polyethylene inserts are at risk for bearing fracture even in the absence of poly wear

Read more

Summary

Conclusions

This case report shows that late bearing in mobile bearing unicompartmental knee arthroplasty can often be a multifactorial event and treatment must address all the risk factors that led to bearing dislocation. Bearing fracture is a very rare complication associated with mobile bearing UKA and patients with thin polyethylene inserts are at risk for bearing fracture even in the absence of poly wear. Keywords: Bearing failure, Mobile bearing, Unicompartmental knee arthroplasty, Bearing dislocation, Bearing fracture, Oxford UKA

Background
Findings
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.