Abstract

We report on online measurements of beam parameters in the soft X-ray and extreme ultraviolet (EUV) spectral range at the free-electron laser FLASH. A compact, self-supporting Hartmann sensor operating in the wavelength range from 6 to 30 nm was used to determine the wavefront quality of individual free-electron laser (FEL) pulses. Beam characterization and alignment of beamline BL1 was performed with λ 13.5 nm /90 accuracy for wavefront rms ( w rms ). A spot size of 159 μm (second moment) and other beam parameters are computed using a spherical reference wavefront generated by a 5 μm pinhole. Beam parameters are also computed relative to a reference wavefront created by a laser-driven plasma source of low coherence, proving the feasibility of such a calibration and reaching λ 13.5 nm /7.5 w rms accuracy. The sensor was used for alignment of the toroidal focusing mirror of beamline BL1, resulting in a reduction of w rms by 25%, and to investigate wavefront distortions induced by thin solid filters.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.