Temporal and dual beam energy subtraction are discussed in conjunction with selective imaging of iodine. Several different kinds of energy subtraction are described and their relative sensitivities to noise are predicted. Effects due to beam hardening are demonstrated for temporal and energy subtraction. Hardening errors are more severe for the latter case by about a factor of three. A possible hardening correction is presented for energy subtraction and an experimental example is provided. Techniques for choosing optimized spectra and data handling for dual beam energy subtraction are presented. These include qualitative principles for spectral selection, identification of the individual noise contributions from the two beams, use of weighted least squares fits, and beam hardening and exposure reduction through the use of filtration.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.