Abstract

We present spin-exchange optical pumping (SEOP) using a third-generation (GEN-3) automated batch-mode clinical-scale 129Xe hyperpolarizer utilizing continuous high-power (~170 W) pump laser irradiation and a novel aluminum jacket design for rapid temperature ramping of xenon-rich gas mixtures (up to 2 atm partial pressure). The aluminum jacket design is capable of heating SEOP cells from ambient temperature (typically 25 °C) to 70 °C (temperature of the SEOP process) in 4 minutes, and perform cooling of the SEOP gas mixture to the temperature at which the hyperpolarized gas mixture can be released from the hyperpolarizer (with negligible quantity of Rb metal leaving the cell) in approximately 4 minutes-substantially faster (by a factor of six) than previous hyperpolarizer designs relying on air heat exchange versus aluminum as employed here. These decreases in temperature cycling time will likely be highly beneficial for the overall increase of production rates of batch-mode (i.e., stopped-flow) 129Xe hyperpolarizers, which is particularly beneficial for clinical applications. The additional advantage of the presented design is significantly improved thermal management of the SEOP cell. Accompanying the heating jacket design and performance, we also evaluate the repeatability of SEOP experiments conducted using this new architecture, and present typically achievable hyperpolarization levels exceeding 40% at exponential build-up rates on the order of 0.1 min-1.