Abstract

We review historical and methodological approaches to measurements of intramuscular pressure (IMP) in humans. These techniques provide valuable measures of muscle tone and activity as well as diagnostic criteria for evaluation of exertional compartment syndrome. Although the wick and slit catheter techniques provide accurate measurements of IMP at rest, their value for exercise studies and diagnosis of exertional compartment syndrome is limited because of low frequency response and hydrostatic (static and inertial) pressure artifacts. Presently, most information on diagnosis of exertional compartment syndromes during dynamic exercise is available using the Myopress catheter. However, future research and clinical diagnosis using IMP can be optimized by the use of a miniature transducer-tipped catheter such as the Millar Mikro-tip.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.