Abstract

The image quality and potential usefulness for patient skin-dose reduction of a newly developed flat-panel detector (FPD) system employing irradiation side sampling (ISS) were investigated and compared to a conventional computed radiography (CR) system. We used the X-ray beam quality of RQA 9 as noted in the standard evaluation method by the International Electrotechnical Commission 62220-1 to evaluate the image quality of the detector for chest radiography. The presampled modulation transfer function (MTF) of the ISS-FPD system was slightly higher than that of the CR system in the horizontal direction at more than 2.2 cycles/mm. However, the presampled MTF of the ISS-FPD system was slightly lower than that of the CR system in the vertical direction. The Wiener spectrum of the ISS-FPD system showed a 50-65 % lesser noise level than that of the CR system under the same exposure condition. The detective quantum efficiency of the ISS-FPD system was at least twice as great as that of the CR system. We conclude that the ISS-FPD system has the potential to reduce the patient skin dose compared to a conventional CR system for chest radiography.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.