Abstract

ObjectiveTo validate a newly designed cochlear implant electrode (TRACE) in the standard monopolar mode and compare it to a patient group implanted with a standard Nucleus Contour™ cochlear implant electrode. The electrode contacts of the TRACE electrode have the same active surface area for stimulation, but the position in the scala tympani is different from that of the Nucleus Contour™ electrode.Material and MethodsThe following parameters, used in cochlear implant fitting and evaluation procedures, were determined: the threshold and comfort stimulation current levels; the electrode impedances; and the phoneme discrimination and speech recognition scores using the ACE™ speech algorithm.ConclusionThe new electrode does not differ significantly from the standard Nucleus Contour™ electrode in terms of the investigated parameters within the test group.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.