Abstract
We adopt the Ensemble Empirical Mode Decomposition (EEMD) method, with an appropriate thresholding on the Intrinsic Mode Functions (IMFs), to denoise the magnetocardiography (MCG) signal. To this end, we discuss the two associated problems that relate to: (i) the amplitude of noise added to the observed signal in the EEMD method with a view to prevent mode mixing and (ii) the effect of direct thresholding that causes discontinuities in the reconstructed denoised signal. We then denoise the MCG signals, having various signal-to-noise ratios, by using this method and compare the results with those obtained by the standard wavelet based denoising method. We also address the problem of eliminating the high frequency baseline drift such as the sudden and discontinuous changes in the baseline of the experimentally measured MCG signal using the EEMD based method. We show that the EEMD method used for denoising and the elimination of baseline drift is superior in performance to other standard methods such as wavelet based techniques and Independent Component Analysis (ICA).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.