Abstract

The Vitamin D Standardization Program (VDSP) coordinated an interlaboratory study to assess the comparability of measurements of total 25-hydroxyvitamin D [25(OH)D] in human serum, which is the primary marker of vitamin D status. A set of 50 individual donor samples were analyzed by 15 different laboratories representing national nutrition surveys, assay manufacturers, and clinical and/or research laboratories to provide results for total 25(OH)D using both immunoassays (IAs) and LC tandem MS (MS/MS). The results were evaluated relative to bias compared with the target values assigned based on a combination of measurements at Ghent University (Belgium) and the U.S. National Institute of Standards and Technology using reference measurement procedures for the determination of 25(OH)D2 and 25(OH)D3. CV and mean bias for each laboratory and assay platform were assessed and compared with previously established VDSP performance criteria, namely CV ≤ 10% and mean bias ≤ 5%. Nearly all LC-MS/MS results achieved VDSP criteria, whereas only 50% of IAs met the criterion for a ≤10% CV and only three of eight IAs achieved the ≤5% bias. These results establish a benchmark for the evaluation of 25(OH)D assay performance and standardization activities in the future.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.