Abstract

In this study, we performed BANG-3® polymer gel dosimetry to propose an experimental technique for relative output factor and 3D dose distribution measurements of small radiosurgical fields.We were irradiated 2 collimators (5 and 30 mm diameter) with 6 MV radiosurgery beam. And, MR scanned with the same slice thickness and three different in plane resolutions. In these experiments, output factor measurements with the Pin Point ion chamber and diode, EBT films are compared with BANG-3® polymer gel. We used a spherical phantom filled with BANG-3® polymer gel to measure 3D dose distributions. The irradiated phantom was scanned with an MRI scanner. Dose distributions were obtained by calibrating the polymer gel for a relationship between the absorbed dose and the spin-spin relaxation rate of the magnetic resistance signal.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.