Abstract

A series of samples of increasing volume (from 0.001 to 4.0 cm 3), containing the same constant concentration (40 g/l) of two simple compounds, ethylbenzoate and 4- tert.-butylphenol were injected on a Kromasil-C 18 column with methanol–water (62:38, v/v) as the mobile phase. Complex band profiles were observed when the volume of the sample became large enough and strong band interference took place. The analysis of the fractions collected during the elution of the mixed band demonstrates that, for samples larger than 2 cm 3, the band of 4- tert.-butylphenol is split into two separate bands, one eluted before and the other eluted after the band of ethylbenzoate. Such a phenomenon has never been observed yet in RPLC, under isocratic elution conditions.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.