Abstract

Balloon rupture during angioplasty can with calcified or recalcitrant lesions. A 61-year-old woman presented with worsening arm and facial swelling. She had a history of left upper extremity thrombolysis and stenting of the innominate vein 6years prior. Venography showed severe in-stent stenosis. After crossing the lesion, a 12-mm balloon was inflated, which ruptured at nominal pressure. The balloon became stuck and could not be moved over the wire even after retraction of the sheath. A limited surgical cutdown was performed, and the balloon and the wire were removed together. The ruptured balloon part was found to be everted and circumferentially wrapped around the wire, preventing the wire exchange. After cutting the everted portion of the balloon, the catheter was removed without losing wire access. A high-pressure balloon was subsequently used to treat the lesion successfully. Her symptoms had resolved on follow-up, and the stent remained patent after 6months.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.