Abstract

We report a case of balloon catheter rupture with subsequent entrapment during percutaneous transluminal coronary angioplasty. The presence of a calcified, distal lesion is believed to have prevented withdrawal of the broken catheter. A nonsurgical retrieval technique, using a second dilation system, was used to free the catheter.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.