Abstract

: Nosocomial infections may result from intensive care unit pulse oximeters. The descriptive study examined pulse oximeter sensor microbiological contamination and the efficacy of manual disinfection with alcohol and sodium hypochlorite in five hospital intensive care units. Sixty-eight reusable pulse oximeter sensors were swabbed, cultured, and evaluated after decontamination. In private and public hospitals, 12 (35.2%) and 13 (37.2%) pulse oximeters tested positive for bacteria. Alcohol 70% reduced the microbial load and more than 10% sodium hypochlorite. The study found that purposeful cleaning and disinfection reduce microorganisms. Alcohol was more efficacious than sodium hypochlorite. Critical care facilities should regularly clean reusable pulse oximeter sensors.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.