Abstract
Hospital-acquired infections (HAIs) are largely preventable through risk analysis and modification of practice. Anaesthetic practice plays a limited role in the prevention of HAIs, although laryngoscope use and decontamination is an area of concern. We aimed to assess the level of microbial contamination of re-usable laryngoscope blades at a public hospital in South Africa. The theatre complex of a secondary-level public hospital in Johannesburg. Blades from two different theatres were sampled twice daily, using a standardised technique, over a 2-week period. Samples were quantitatively assessed for microbial contamination, and stratified by area on blade, theatre and time using Fisher's exact test. A contamination rate of 57.3% (63/110) was found, with high-level contamination accounting for 22.2% of these. Common commensals were the most frequently isolated micro-organisms (79.1%), but important hospital pathogens such as Enterobacter species and Acinetobacter baumannii were isolated from blades with high-level contamination. No significant difference in the level of microbial contamination by area on blade, theatre or time was found (p<0.05). A combination of sub-optimal decontamination and improper handling of laryngoscopes after decontamination results in significant microbial contamination of re-usable laryngoscope blades. There is an urgent need to review protocols and policies surrounding the use of these blades.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.