Abstract
The in vitro adherence of Staphylococcus aureus, Staphylococcus epidermidis and Escherichia coli to five commercially available prosthetic vascular graft materials was compared. The influence of precoating the segments with human plasma for 2 h was also studied. S35-methionine was used to radiolabel bacteria. The segments were exposed to bacterial suspensions of approximately 10(7) CFU/ml at 37 degrees C for 0.5-18h. Following repeated washing in phosphate buffered saline (PBS), radioactivity associated with the segments was measured. The adherence of the three clinically relevant bacterial species was higher to untreated Dacron than to gelatin or collagen impregnated/coated Dacron or to PTFE. Furthermore, precoating of grafts with human plasma reduced bacterial adherence to woven Dacron, had a little effect on gelatin coated Dacron, but increased the adherence to collagen treated Dacron and, in particular, to PTFE.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.