Abstract

Simulated continuous arteriovenous hemodialysis (CAVHD) was performed with Renaflo anisotropic polysulfone capillary hemofilters and Biospal homogeneous PAN parallel-plate hemofilters. Inulin and glucose were present in the dialysis solution. Plasma and dialysis solution flow rates were varied as was the delivery of dialysis solution by gravity or pump. The presence of dialysis solution did not affect the components of transmembrane pressure and did not alter ultrafiltration rate. Inulin and glucose backtransported into the plasma perfusate during CAVHD with the Biospal but not with the Renaflo filters. Dialysis solution will need to be sterile and pyrogen-free even in the low-flow, low-pressure system of CAVHD.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.