Abstract

The molecular hydride negative ions of mass 129 amu, 127IH2−, 127ID− and 128TeH−, have been studied in order better to understand potential backgrounds in AMS measurements of 129I. All three of these ions appear to be stable or, if metastable, to have lifetimes that are sufficiently long to allow at least some of them to reach the high voltage terminal of the accelerator. It is found, however, that these make only a minor contribution to the combined rate of 127I7+ and 128Te7+ ions that pass around the analysing magnet when it is set for 129I7+ detection. The principal contribution comes instead from the injection of 127I− ions into the accelerator along with the 129I− ions, despite the use of a low Cs sputtering energy in the ion source that was intended to eliminate them. This contribution is, however, small and is effectively removed by a Wien filter. In addition, a sample of Woodward iodine, obtained from the University of Arizona, was found to have an 129I/127I ratio of (13.3 ± 0.9) × 10−15. This is the lowest ratio so far reported. In contrast, the batch of Woodward iodine that was being used as an iodine carrier in our laboratory was found to have a much higher ratio of (78.2 ± 2.4) × 10−15. A search for a lower ratio blank failed to find a suitable material.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.