Abstract

We have developed a technique to accurately null the longitudinal magnetization (Mz) of background material. This suppression involves first saturating the longitudinal magnetization (Mz) of a region, and then applying several nonselective inversions. The inversions are timed relative to the saturation such that Mz is nulled across a broad range of T1 at a predetermined time after the initial saturation. B1 and B0 inhomogeneity, which could lead to inaccurate suppression, are dealt with by the combination of a multiple tip saturation sequence and four adiabatic inversion pulses. The suppression sequence can be used to form projective angiograms by selectively tagging the imaging region with the saturation pulse. After the inversions are played out, a projection taken through the tag region when Mz is nulled will only contain signal from blood that has flown into the region after the saturation. Since only two dimensions are acquired, the technique can acquire gated projection angiograms in reasonable scan times. Representative inflow MIR angiograms of the carotid arteries and renal arteries show excellent background suppression.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.