Abstract

To assess the difference in intraocular tumors height measurements intraoperatively before and after radioactive plaque application. Twenty-four eyes of 24 patients with intraocular tumors, candidates for radioactive therapy, were included. Each tumor was measured at the same setting before and after plaque application independently by 3 sonographers, using a 20-MHz B-scan transducer. Mean pre-plaque and post-plaque measurements were calculated and recorded. An A-scan vector aided in identification of the inner and outer tumor surfaces. Each patient was examined independently three times by three experienced ultrasonographers within the same setting to assess interobserver variability. There was no statistically significant difference between the 3 examiners' readings, with P-value 0.99 for pre-plaque height and 0.97 for post-plaque height. Mean pre-plaque height was 5.16±2.11mm, while post-plaque height was 5.51±2.1mm (P-value 0.001). The Spearman correlation test showed that initial tumor height was negatively correlated with the difference between both heights, but with no statistical significance. Intraocular tumor height measurement differs significantly before and after plaque application. Use of the pre-plaque height is advised until further studies are performed to assess the effect of this difference on treatment outcomes.