Abstract

In this study it was evaluated the variation of tin concentration in function of storage time in canned food samples by using an axial view inductively coupled plasma optical emission spectrometry (AX ICP OES). The effect of various parameters of the ICP OES, such as analytical lines, sample introduction systems and interferences were investigated. Eight emission lines for tin were tested, but only 283.998nm line was found to be robust enough with respect to interferences and sensitivity, presenting a limit of quantification of 0.094mgkg−1. A concentric nebuliser combined with a cyclonic chamber has shown better sensitivity than a V-groove nebuliser. Tin was determined in canned tomato sauce stored at room temperature (20.0±1.8°C) for 0, 30, 60, 90, 120 and 150days. After 150days, tin contents increased in the 43–91% range. Despite an increase of almost 100% for one sample, tin contents are well below the maximum level allowed by Agência Nacional de Vigilância Sanitária (ANVISA) regulation of 250mgkg−1.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.