Abstract

Calibration of axial quadrant photodetector (QPD) signal to the trapped bead position in an optical tweezer is important to measure the quantitative mechanical parameter in axial (laser propagation) direction. An alternative calibration based on the Stuck Bead Method (SBM) was proposed in this study. 3 µm polystyrene beads were stuck at the surface of glass coverslip and moved axially around the laser focus. QPD was used to obtain the position dependent intensity profile at three different laser powers (19.8 mW, 34.1 mW, 48.5 mW). The QPD signal-to-distance calibration value was consistent at 26 mV/µm for the used bead at the three laser powers. It was found that the calibration values are independent of laser powers and limited by the resolution of distance adjustment.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.