Abstract

Approximately 250 computed tomography (CT) systems were surveyed in a nationwide study to determine the average radiation dose resulting from a typical adult head procedure. The multiple scan average dose (MSAD) was selected as the dose descriptor. For the typical adult CT head procedure, the MSAD was generally within 2.2-6.8 rads (22-68 mGy). Variations in dose by a factor of two or more were often seen for a given manufacturer and model. These dose ranges indicate a potential to reduce dose by carefully selecting imaging techniques. Overall, variations in dose can result from differences in the user's choice of technique (desired image quality) or from actual differences in scanner performance (caused by differences in collimation, filtration, or geometry). To use CT appropriately, a facility should consider dose as well as image quality in selecting optimal techniques for typical modes of operation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.