Abstract
The most important reference for breast cancer diagnosis is still the mammography. Mammographic technology has evolved in such a way that we have digital systems nowadays. Thus, image generation quality control must evolve as well. In this work, we have evaluated digital image quality and entrance skin dose (ESD) for different breast thicknesses simulated by PMMA plates (one, two, three and four centimeters) for different target filter combinations, suggesting a protocol optimization to the mammographic automatic system.We have used CDMAM 3.4 phantom (a contrast detail phantom) and its software in order to analyze digital images from a Siemens Mammomat Inspiration mammographic equipment. It has been used three combinations of target-filter: Molybdenum Molybdenum, Molybdenum Rhodium, Tungsten Rhodium. Different peak tensions were tested to each configuration: 24 kVp, 26 kVp, 28 kVp and 30 kVp. Altogether 52 different configurations were tested. In order to measure ESD, we used Radcal’s 10x5–6M ionization chamber.Results show that automatic configuration does not offer an optimized process. Tungsten Rhodium was the best option for all breast thicknesses, once it offered the best relation between image quality and mean glandular dose. This result differs from literature which states that the Molybdenum Molybdenum combination is the best option for thin breasts.
Highlights
The most important reference for breast cancer diagnosis is still the mammography
Mammographic technology has evolved in such a way that we have digital systems nowadays
We have evaluated digital image quality and entrance skin dose (ESD) for different breast thicknesses simulated by PMMA plates for different target filter combinations, suggesting a protocol optimization to the mammographic automatic system.We have used CDMAM 3.4 phantom and its software in order to analyze digital images from a Siemens Mammomat Inspiration mammographic equipment
Summary
O câncer de mama é a doença neoplásica com maior taxa de mortalidade entre mulheres no Brasil. Em um equipamento de mamografia por raios X, os materiais do ânodo e do filtro determinam a forma do espectro gerado pelo tubo. Com o avanço da tecnologia digital e da automatização dos sistemas em radiodiagnóstico, se faz necessário avaliar se o funcionamento dos mamógrafos estão ocorrendo de forma otimizada, isto é, proporcionando a melhor qualidade de imagem com a menor dose glandular média (DGM) possível. Analisouse se as combinações ânodo-filtro utilizadas pelo mamógrafo eram as mais adequadas para diferentes espessuras de mama e verificou-se, também, se o mamógrafo avaliado trabalha com seu controle automático de exposição otimizado (OPDOSE), isto é, se a seleção de tensão de pico e carga transportável fornece a imagem com a melhor qualidade e com a menor dose possível
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.