Abstract

We report on avalanche-ion back-flow measurements in the novel Micro-Hole and Strip-Plate (MHSP) multiplier and in gaseous photomultipliers comprising Gas Electron Multipliers (GEMs) followed by an MHSP. In a 3-GEMs/MHSP photomultiplier with reflective photocathode, avalanche-ion back-flow fraction of ∼7% and ∼2% were recorded for respective effective gains of 10 7 and 10 6, in Ar/CH 4 (95/5) at 760 Torr. This is about one order of magnitude reduction in ion back-flow compared to the best values measured in 4-GEMs photomultiplier at the same gain. We describe the mode of operation of the MHSP and explain its ion back-flow reduction features.