Abstract

Recently a new fast magnetic resonance imaging strategy, SMASH, has been described, which is based on partially parallel imaging with radiofrequency coil arrays. In this paper, an internal sensitivity calibration technique for the SMASH imaging method using self-calibration signals is described. Coil sensitivity information required for SMASH imaging is obtained during the actual scan using correlations between undersampled SMASH signal data and additionally sampled calibration signals with appropriate offsets ink-space. The advantages of this sensitivity reference method are that no extra coil array sensitivity maps have to be acquired and that it provides coil sensitivity information in areas of highly non-uniform spin-density. This auto-calibrating approach can be easily implemented with only a small sacrifice of the overall time savings afforded by SMASH imaging. The results obtained from phantom imaging experiments and from cardiac studies in nine volunteers indicate that the self-calibrating approach is an effective method to increase the potential and the flexibility of rapid imaging with SMASH.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.