Abstract

This paper describes the automation of an imaging internal source laser desorption Fourier transform mass spectrometer (I2LD-FTMS). The I2LD-FTMS consists of a laser-scanning device [Scott and Tremblay, Rev. Sci. Instrum. 2002, 73, 1108–1116] that has been integrated with a laboratory-built FTMS using a commercial data acquisition system (ThermoFinnigan FT/MS, Bremen, Germany). A new user interface has been developed in National Instrument's (Austin, Texas) graphical programming language LabVIEW to control the motors of the laser positioning system and the commercial FTMS data acquisition system. A feature of the FTMS software that allows the user to write macros in a scripting language is used creatively to our advantage in creating a mechanism to control the FTMS from outside its graphical user interface. The new user interface also allows the user to configure target locations. Automation of the data analysis along with data display using commercial graphing software is also described.