Abstract
In many IBA applications the aim is to obtain quantitative figures characterising the sample, usually requiring a reliable measurement of the beam current for normalisation. To avoid the drawbacks associated with on-sample and post-sample charge measurement systems a pre-sample measurement system can be used.In this work an upgrade to the pre-sample charge measurement system at the Lund Ion-beam Analysis Facility is presented, in which a custom built control module based around a Teensy 3.2 micro-controller has been added to the existing system. The new control module allows for remote handling and monitoring of the deflection system, operated via USB connection to a PC running a graphical user interface. Details of the control modules functionality, design and programming are given, as well as the functionality of the user interface. An example of how the new control module can be used for automated calibration of the deflection system is also presented.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.