Abstract

A simple modification to an existing adult ventilator has been designed to permit mechanical ventilation of neonates and small children with the Ayre's T-piece circuit. The Nuffield Anaesthesia Ventilator Series 200 has been modified by replacing the piston in the patient valve with a fixed leak. This arrangement avoids the problems of critical adjustment commonly encountered with other methods of converting adult ventilators to paediatric use, and allows the ventilator controls to be used normally. The ventilator has been assessed according to the proposed International Standards Organization specifications for the evaluation of the performance of lung ventilators, and the results are described in detail. The modified ventilator is shown to perform as a time-cycled pressure generator capable of delivering tidal volumes between 10 and 300 ml at frequencies from 10 to 85/minute. It is therefore ideally suited to neonatal and paediatric use.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.