Abstract

PurposeTo construct a method and software to track gold seed implants in prostate and lung patients undergoing radiotherapy using CBCT image projections. MethodsA mathematical model was developed in the MatLab (Mathworks, Natick, USA) environment which uses a combination of discreet cosine transforms and filtering to enhance several edge detection methods for identifying and tracking gold seed fiducial markers in images obtained from Varian (Varian Medical Systems, Palo Alto, USA) and Elekta (Kungstensgatan, Sweden) CBCT projections. ResultsOrgan motion was captured for 16 prostate patients and 1 lung patient. ConclusionImage enhancement and edge detection is capable of automatically tracking markers for up to 98% (Varian) and 79% (Elekta) of CBCT projections for prostate and lung markers however inclusion of excessive bony anatomy (LT and RT LAT) inhibit the ability of the model to accurate determine marker location.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.