Background and purposeAnatomical changes during external beam radiotherapy prevent the accurate delivery of the intended dose distribution. Resolving the delivered dose, which is currently unknown, is crucial to link radiotherapy doses to clinical outcomes and ultimately improve the standard of care. Material and methodsIn this study, we present a dose reconstruction workflow based on data routinely acquired during MR-guided radiotherapy. It employs 3D MR images, 2D cine MR images and treatment machine log files to calculate the delivered dose taking intrafractional motion into account. The developed pipeline was used to measure anatomical changes and assess their dosimetric impact in 89 prostate radiotherapy fractions delivered with a 1.5 T MR-linac at our institute. ResultsOver the course of radiation delivery, the CTV shifted 0.6 mm ± 2.1 mm posteriorly and 1.3 mm ± 1.5 mm inferiorly. When extrapolating the dose changes in each case to 20 fractions, the mean clinical target volume D98% and clinical target volume D50% dose-volume metrics decreased by 1.1 Gy ± 1.6 Gy and 0.1 Gy ± 0.2 Gy, respectively. Bladder D3% did not change (0.0 Gy ± 1.2 Gy), while rectum D3% decreased by 1.0 Gy ± 2.0 Gy. Although anatomical changes and their dosimetric impact were small in the majority of cases, large intrafractional motion caused the delivered dose to substantially deviate from the intended plan in some fractions. ConclusionsThe presented end-to-end workflow is able to reliably, non-invasively and automatically reconstruct the delivered prostate radiotherapy dose by processing MR-linac treatment log files and online MR images. In the future, we envision this workflow to be adapted to other cancer sites and ultimately to enter widespread clinical use.

Highlights

  • Background and purposeAnatomical changes during external beam radiotherapy prevent the accurate delivery of the intended dose distribution

  • We have presented an end-to-end pipeline to harness the on-board imaging capabilities of an MR-linac to reconstruct the delivered radiotherapy dose

  • We found the observed anatomical changes and their effect on the dose distribution to be small and most likely clinically irrelevant

Read more

Summary

Introduction

Background and purposeAnatomical changes during external beam radiotherapy prevent the accurate delivery of the intended dose distribution. Material and methods: In this study, we present a dose reconstruction workflow based on data routinely acquired during MR-guided radiotherapy. It employs 3D MR images, 2D cine MR images and treatment machine log files to calculate the delivered dose taking intrafractional motion into account. The developed pipeline was used to measure anatomical changes and assess their dosimetric impact in 89 prostate radiotherapy fractions delivered with a 1.5 T MR-linac at our institute. Bladder D3% did not change (0.0 Gy ± 1.2 Gy), while rectum D3% decreased by 1.0 Gy ± 2.0 Gy. anatomical changes and their dosimetric impact were small in the majority of cases, large intrafractional motion caused the delivered dose to substantially deviate from the intended plan in some fractions.

Objectives
Methods
Results
Conclusion