Abstract

Transorbital sonography provides reliable information about the estimation of intra-cranial pressure by measuring the optic nerve sheath diameter (ONSD), whereas the optic nerve (ON) diameter (OND) may reveal ON atrophy in patients with multiple sclerosis. Here, an AUTomatic Optic Nerve MeAsurement (AUTONoMA) system for OND and ONSD assessment in ultrasound B-mode images based on deformable models is presented. The automated measurements were compared with manual ones obtained by two operators, with no significant differences. AUTONoMA correctly segmented the ON and its sheath in 71 out of 75 images. The mean error compared with the expert operator was 0.06 ± 0.52 mm and 0.06 ± 0.35 mm for the ONSD and OND, respectively. The agreement between operators and AUTONoMA was good and a positive correlation was found between the readers and the algorithm with errors comparable with the inter-operator variability. The AUTONoMA system may allow for standardization of OND and ONSD measurements, reducing manual evaluation variability.