Abstract
Of 115 patients with chronic open angle glaucoma or suspected glaucoma 115 eyes were subjected to a visual field examination with the manual Goldmann perimeter and the automatic Octopus perimeter. In 84.4 +/- 8.9% the automaton detected more visual field loss, including 27.8 +/- 10.9% of the eyes where no visual field loss was found by the manual perimetry. Since some eyes with normal fields on manual perimetry and abnormal fields detected by automatic perimetry remained abnormal on retesting, it is assumed that scotomas found by automatic perimetry and not shown by manual perimetry constitute false-negative manual fields rather than false-positive automatic fields.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie. Albrecht von Graefe's archive for clinical and experimental ophthalmology
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.