Abstract
An automatic system for measuring micro flow rate of liquid is presented. A computer mouse device was modified as an optical encoder for this purpose. All the sensing and electrical signal are processed by the mouse device. A very simple structure has been realized using the PS/2 mouse interface to get electric power and transfer the data to computer directly. The data treatment was programmed using Microsoft Visual Basic. This system shows a useful solution for micropump evaluation. Because this device works in digital On/Off principle, there is no lower limit for measurable range of flow rate and bidirectional measurement is also possible. The full measurable range is up to 115 /spl mu/l/min, at which the measurement precision is assured within /spl plusmn/2%. The resolution is dependent on the sampling speed of the computer (0.05 second). The device is not affected by the change of temperature or thermal capacity of fluids. It is an alternate as a tool to calibrate mass flow meter. Automatic parallel measurements in multi-channel are also possible.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.