Abstract

The pulsed electro-acoustic method (PEA) is widely used both in laboratory for fundamental research and in industry for high-voltage direct current cable assessment and diagnosis. It is well known that the electrical stress in dielectrics can be influenced by space charges. In this study, we have modified and upgraded the PEA system for better resolution, safety, and measurement of the space charge distributions automatically in a solid dielectric during the application of various DC high voltages. We adapt the PCI extensions for instrumentation system with a DSO/DAQ system for automatic measurements. The system could be used to automatically measure the space charge distribution with higher accuracy under various applied DC voltages.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.