Midline shift (MLS) is an important quantitative feature clinicians use to evaluate the severity of brain compression by various pathologies. The midline consists of many anatomical structures including the septum pellucidum (SP), a thin membrane between the frontal horns (FH) of the lateral ventricles. We proposed a procedure that can measure MLS by recognizing the SP within the given CT study. The FH region is selected from all ventricular regions by expert rules and the multiresolution binary level set method. The SP is recognized using Hough transform, weighted by repeated morphological erosion. Our system is tested on images from 80 patients admitted to the neurosurgical intensive care unit. The results are evaluated by human experts. The mean difference between automatic and manual MLS measurements is 0.23±0.52mm. Our method is robust and can be applied in emergency and routine settings.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.